Description
Specification Name Detail
SKU HI2030-01
pH Range pH: -2.000 to 16.000, -2.00 to 16.00; ±1000 mV
pH Resolution 0.001 pH, 0.01 pH; 0.1 mV
pH Accuracy (@25°C/77°F) ±0.01 pH, ±0.002 pH; ±0.2 mV
pH Calibration Automatic, up to 3 points (5 points*) calibration, 5 standard (7 standard*) buffers available (1.68*, 4.01 or 3.00, 6.86, 7.01, 9.18, 10.01, 12.45*) and 2 custom buffers*
pH Temperature Compensation Automatic -5 to 100˚ C (23 to 212˚ F) (using integral temperature sensor)
EC Range 0.00 to 29.99 μS/cm, 3.00 to 29.99 μS/cm, 30.0 to 299.9 μS/cm, 300 to 2999 μS/cm, 3.00 to 29.99 mS/cm, 30.0 to 200.0 mS/cm, up to 500.0 mS/cm
EC Resolution 0.01 μS/cm, 00.1 μS/cm, 0.01 μS/cm, 0.01 mS/cm, 00.1 mS/cm
EC Accuracy ±1% of reading (±0.05 μS/cm or 1 digit, whichever is greater)
EC Calibration 1 Cell Factor calibration; 6 standards available: 84, 1413 μS/cm, 5.00, 12.88, 80.0, 111.8 mS/cm, 1 point offset: 0.00 μS/cm
TDS Range 0.00 to 14.99 ppm (mg/L), 15.0 to 149.9 ppm (mg/L), 150. to 1499. ppm (mg/L), 1.50 to 14.99 g/L, 15.0 to 100.0 g/L, up to 400.0 g/L absolute TDS ‡ (with 0.80 factor)
TDS Resolution 0.01 ppm, 0.1 ppm, 1 ppm, 0.01 g/L, 0.1 g/L
TDS Accuracy ±1% of reading (±0.03 ppm or 1 digit, whichever is greater)
TDS Calibration through EC calibration
EC/TDS Temperature Compensation Automatic -5 to 100 ºC (23 to 212 ºF) NoTC – can be selected to measure absolute conductivity.
Temperature Correction Coefficient 0.00 to 6.00% / ºC (for EC and TDS only). Default value is 1.90% / ºC
EC to TDS Conversion Factor 0.40 to 0.80 (default value is 0.50)
Salinity Range 0.0 to 400.0% NaCl, 0.01 to 42.00 PSU, 0.0 to 80.0 g/L*
Salinity Resolution 0.1% NaCl, 0.01 PSU, 0.01 g/L
Salinity Accuracy ±1% of reading
Salinity Calibration 1 point with HI 7037L 100% NaCl sea water standard (other scales through EC calibration)
Salinity Temperature Compensation 0 to 40 g/L (with 1 g/L resolution)
Sodium Chloride Range 0.0 to 400.0% NaCl
Sodium Chloride Resolution 0.1% NaCl
Sodium Chloride Accuracy ±1% of reading
Sodium Chloride Calibration 1 point with HI 7037L 100% NaCl sea water standard (other scales through EC calibration)
DO Range 0.00 to 45.00 ppm (mg/L) 0.0 to 300.0% 0.0 to 50.0 °C (32.0 to 122.0 °F)**
DO Accuracy (@25˚C/77˚F) ±1.5% of reading or ±1 digit
DO Resolution 0.01 ppm (mg/L); 0.1% saturation
DO Calibration One or two points at 0% (HI 7040) and 100% (water saturated air)
DO Temperature Compensation 0.0 to 50.0 ˚C (32.0 to 122 ˚F)
Salinity Compensation 0 to 40 g/L, 1 g/L
DO Altitude Compensation -500 to 4000 m (with 100 m resolution)
Temperature Range -20.0 to 120.0 ˚C, -4.0 to 248.0 °F
Temperature Resolution 0.1 °C, 0.1 °F
Temperature Accuracy ±0.5 °C, ±1.0 °F
pH Electrode HI11310 digital double junction, refillable with 3.5M KCl electrolyte solution pH electrode, built in temperature sensor, and 3.5 mm connector.ᵻ
EC/TDS Probe HI763100 EC/temperature probe
DO Probe HI764080 DO/temperature probeᵻ
Power Supply 5 VDC adapter (included)
Environment 0 to 50°C (32 to 122°F), RH max 95% non-condensing
Dimensions 202 x 140 x 12.7mm (8” x 5.5” x 0.5”)
Weight 250 g (8.82 oz)
Ordering Information HI2030 EC kit includes: HI763100 Conductivity probe, 4 sachets of 1413 μS/cm conductivity standard, 2 sachets of 5000 μS/cm conductivity standard, 2 sachets of 12880 μS/cm conductivity standard, 2 sachets general purpose rinse solution, USB cable, benchtop docking station with electrode holder, wall mount cradle, 5 VDC power adapter, quality certificate and instruction manual.
Special Notes ‡ absolute conductivity *temperature limits will be reduced to actual probe/sensor limits ** with temperature compensation function disabled ᵻAvailable for purchase separately
Warranty 2 years
Reviews